The ALTA 3-Part Hematology Analyzer (ADX-HEME-340) uses Electrical Impedance and Photometric (for HGB measurement) principles for detecting whole blood parameters like RBC, WBC, PLT, etc., for Complete Blood Count (CBC).
Features of ALTA Hematology Analyzer:
- Throughput up to 80 samples/hour with whole blood mode
- 10.4-inch color display screen with touch panel
- Cell pulse baseline tracking and restore technology
- High voltage burning technology
- Built-in liquid sensor
- High-speed pulse collecting and discrimination
- Intelligent floating threshold technology
- Large storage capacity
- 500,000 results including histogram & Patient information
ALTA Hematology Analyzer – For those who never compromise on quality:
LOW SAMPLE VOLUME:
- Venous & Capillary whole blood mode: 9.6 µL per test
- Prediluted blood mode: 20 µL per test
TEST THROUGHPUT & PARAMETERS /LARGE MEMORY:
- 60 Tests/hour
- 21 parameters and 3 histograms
- Customized parameter printing with flexible report formats
- 2,00,000 samples of data storage
ACCURATE, STABLE AND RELIABLE:
- Cell particles guide, accurate signals to reflect the size of cells
- PLT simulation curve, reduce the influence of RBC on PLT counting
- External wide power supply & anti-interference components make the instrument highly stable
COMPLETE CALIBRATION AND QC SYSTEM:
- Manual & automatic calibration
- L-J & X-B QC facility
EASY MAINTENANCE:
- 3-door design, convenient for maintenance
- Self-test function, automatic pipeline washing during start-up and shut down
- Built-in Clog prevention mechanism for smooth flow
Technical Specifications of ALTA Hematology Analyzer:
Testing principle: Electrical impedance, colorimetric method for HGB with cyanide-free hemolytic agent
Display: 10-inch colour LED touchscreen display, supports external mouse and keyboard
Parameters: WBC, LYM#, LYM%, MID#, MID%, GRA#, GRA%, RBC, HGB, HCT, MCV, MCH, MCHC, PLT, RDW-CV, RDW-SD, MPV, PCT, PDW, P-LCR, P-LCC, RBC histogram, WBC histogram and PLT histogram
Data storage: Storing over 2,00,000 samples (histogram, patient information), multiple search modes as per requirements, combined preview for output data
Print mode: Build-in printer, optional connection with an external printer
Interface: USB×4, RS-232×1, Ethernet×1
Power supply: Voltage : A.C 100V~240V, Input power : ≤150VA, Frequency : 50/60 Hz
Reviews